◇◇HOME◇◇
 
 
 
 
 
 
 
 
 
EQUIPMENT & SYSTEMS FOR RESEARCH

We design and build almost any sort of equipment or systems based on our
customers' specification using our proven technology and knowledge.
As follows:

    High temperature & pressure autoclaves.(with magnetic stirrer)
    Synthetic resin research equipment.(with direct drive stirrer)
    High temperature & pressure test apparatus.
    High temperature & pressure corrosion test apparatus.
    Multiple-purpose cracking mini-plants.
    Catalyst evaluation & reforming apparatus
    Hydrogenation, hydrodesulfurization test apparatus.
    Gas-phase oxidation reactors.
    etc.


(AUTOCLAVES)

( Autoclave with jacket and magnetic stirre r)

This autoclave is to be heated by steam or hot oil.
Magnetic stirrer is installed.

Design Pressure : 10 MPa
Design Temperature : 200 C
Volume : 10 L
Main material : SUS316
Meets all requirements of the High Pressure Gas
Safety Law (HPGSL). (Toxic)

 


( Autoclave with electric furnace and magnetic stirrer )

This autoclave is to be heated by an electricheater.
Magnetic stirrer is installed and a control panel is supplied.

Design Pressure : 40 MPa
Design Temperature : 300 C
Volume : 3 L
Main material : SUS316
Approved by HPGSL.(Toxic)

 


(AUTOCLAVE UNIT)

Although our customers have various requirements for structure, materials and condition for use, we have established the "TAU series" as follows. We believe that this series will meet our customer's requirements almost perfectly. We can supply autoclaves which conform to the HPGSL.
We have also provided the control panel for maintaining a stable electric furnace temperature.
Model No. TAU-0.1 TAU-0.3 TAU-0.5
Volume
100 mL
300 mL
500 mL
D.P.(bar)
50 100 300
50 100 300
50 100 300
D.T.(℃)
300 300 300
300 300 300
300 300 300
Material
SUS316
SUS316
SUS316
Magnetic stirrer Torque(kg-cm)
3.0
3.0
3.0

Model No. TAU-1 TAU-3 TAU-5
Volume
1.0 L
3.0 L
5.0 L
D.P.(bar)
50 100 300
50 100 300
50 100 300
D.T.(℃)
300 300 300
300 300 300
300 300 300
Material
SUS316
SUS316
SUS316
Magnetic stirrer Torque(kg-cm)
5.0
5.0
5.0

Remarks :

The maximum operating pressure should be 85% of the design one. and the maximum operating temperature should be the design one.

The standard maximun revolution of the magnetic stirrer is 700 rpm.

Bigger torque magnetic stirrer is available.

Other material or design condition are available depending on your request, please inquire to us.


FLOW-THROUGH TYPE APPARATUS
 
( Methanol reformer apparatus )

This apparatus is for evaluation of catalyst for methanol reforming.

Design Pressure : 1.0 MPa
Design Temperature : 300 C
Reactor Volume : 3 L
Main material : SUS316

 


( Alkylization test apparatus (4 lines) )

This apparatus is for testing catalyst for alkylizing
gasoline. Each lines is independent.

Fluid : Gasoline
Design pressure : 1.0 MPa
Design temperature : 450 c
Main material : SUS316


( Flow-through reaction test apparatus (2 lines) )

This apparatus is for evaluation of catalyst's durability in a
reaction under certain conditions. Each line is independent.

Design Pressure : 4.0 MPa
Design Temperature : 350 C
Main material : SUS316

 


( Methanol cracking test plant )

This bench-scale plant is for testing methanol
gasification and separation. This has two reactors.

Fluid : Methanol
Design pressure : 3.0 MPa
Design temperature : 400 c
Main material : SUS316
Law :HPGSL(Toxic)

 


We have a wealth of know-how about designing equipment or plants.
Please don't hesitate to contact us.
 
Our person in charge : Yoji Noda
Tel : +81-82-237-6255 Fax : +81-82-230-0611
E-mail :info@toyokoatsu.co.jp